Death by Lethal Injection: Missouri

views updated

Death by Lethal Injection: Missouri

Book excerpt

By: Fred A. Leuchter

Date: October 15, 1988

Source: Fred A. Leuchter. "Death by Lethal Injection: Missouri." In Lethal Injection Machine Manual: State of Missouri. Boston, Mass.: Fred A. Leuchter Associates, 1988.

About the Author: Fred A. Leuchter ran a company which specialized in the design and manufacture of execution equipment, including devices for delivering execution and lethal injection. Leuchter famously argued that the Holocaust could not have taken place after traveling to concentration camps to carry out research into gas chambers. In 1990, he was exposed as having no formal engineering qualifications, even though he sold his equipment to several states, including Missouri, as detailed below.

INTRODUCTION

Lethal injection was introduced into the United States as a means of execution when it was adopted by the state of Oklahoma in 1977. The first person to be executed by lethal injection was Charles Brooks, in Texas, in 1982. Since then, 842 people have been executed by lethal injection in the United States. It has become the leading form of execution, used by thirty-seven of the thirty-eight states that have the death penalty. The concept of execution by lethal injection arose after a number of botched hangings in New York in the late 19th century. A committee was set up to consider alternatives to hanging and came out in favor of electrocution, as lethal injection was strongly disapproved of by the medical profession. The subject came up again at the start of the modern era of the death penalty, when it was ruled constitutional again in 1976 by the Supreme Court. By that time, modern medical technology made the idea of lethal injection more of a practical proposition.

Death by lethal injection involves three different drugs. The prisoner is strapped down to a gurney and two needles attached to long tubes are inserted into veins in the arms. The tubes are connected to several intravenous drips though a hole in a cement block wall. One drip contains saline and this is started first. At a signal from the warder, a curtain is raised so the prisoner is exposed to witnesses in an adjoining room. Then a drip with an anesthetic, sodium thiopental, is started and the prisoner is put to sleep. Next, a drug called pancuronium bromide (Pavulon) is injected. This paralyzes all the muscles so the prisoner stops breathing. Finally, potassium chloride is injected and this stops the heart. Therefore, the condemned prisoner dies from a combination of anaesthetic overdose and respiratory and cardiac arrest. Sodium thiopental and pancuronium bromide are both used clinically, but the doses used in lethal injection are far higher than the therapeutic doses. The manual extract below describes the lethal injection process in more detail.

PRIMARY SOURCE

EXECUTION BY LETHAL INJECTION: MISSOURI

The Fred A. Leuchter Associates, Inc. Modular Lethal Injection Machine is designed to sequentially deliver variable quantities of three solutions at variable time periods. Utilizing standard 60cc disposable syringes, it is essentially driven by weighted pistons depressing the syringe plungers. It is designed to sequentially deliver the solutions at time intervals controlled by an electrical, solid-state, timing system powered by a sealed battery which receives its charge from the power line. The system is designed to be activated by two executioners: a solid-state digital circuit will randomly determine which executioner controls the system, but will not retain the chosen executioner in memory. Since it operates from battery power, it is not subject to power line failure. Three electrical solenoids cause the weighted pistons to depress the respective syringe plungers. Chemical delivery is via a single intravenous line, commoned by an eight-port stainless steel manifold with luer lock entry ports. This line joins with a standard intravenous administration set dispensing saline solution on a continuous basis into the subject. Three solid state timers operate the solenoids and the stages are initiated by switch-controlled relays monitoring the volume of solution in each syringe.

In the event of a timing system failure, an electrical override is provided to control the solenoid operation. Two switches are used to control the depression of each successive plunger. Any one, or all three plungers, may be controlled, depending upon the problem encountered. The controlling switch group is preset during machine makeready. Thus, the system still functions as a one-of-two executioner controlled system. In the event of total electrical failure, or a partial mechanical failure of the Prime System, a second Back-up System, which is strictly mechanical, is employed. This system consists of a redundant set of piston driven syringes and is operated by three sets of double pull knobs. Determination of which executioner will control the system is made during makeready. Both knobs of each set are pulled in tandem sequentially and the system operates as a one-of-two executioner system. The use of either back-up system requires that sequential timing between successive plunger actuations be determined by clock and command. All three systems have been engineered with a minimal amount of components for trouble free operation, and it is unlikely that a systems failure should occur in any of the three systems. However, should failure occur, the redundant systems should insure a problem-free execution.

Provision is made for systems purge via two saline filled syringes and an external saline supply is utilized to prevent coagulation at the needle tip, in the usual manner. The system is designed to accept eight 60cc syringes; two for saline, two for Pentathol, two for Pancuronium Bromide, and two for Potassium Chloride. Each of the six functional syringes has a positive piston stop pin to prevent premature operation of the system.

The system consists of two modules; The Control Module and The Delivery Module. The Control Module is made up of the power, sequential timing and control circuitry. The Delivery Module consists of the electromechanical and mechanical drive assemblies, syringes, and manifold, to ensure proper fluid delivery.

THE CONTROL MODULE

The Control Module is designed utilizing state-of-the-art, solid-state, power, digital and timing circuitry and conventional electrical switches. The on/off and function switches are key controlled to prevent accidental operation. The battery, power supply and charging circuit are also part of this package.

The Control Panel itself contains the Key-controlled on/off switch and two executioner switches (which are pushed simultaneously) for the prime system; the key-controlled function switch for the electrical back-up system (to determine which executioner activates the system) and six operating switches, arranged in three pairs, for the back-up system. These switches are thrown simultaneously by the executioners and control the sequential dispensing of the fluids, individually. Each pair is thrown simultaneously with the proper time interval between operations when the electrical back-up system is utilized.

There are two sets of monitoring lights, one on The Control Module and a redundant set on The Delivery Module. A system-on light indicates power on. Each syringe is monitored by a three light sequence. Red indicates Ready, Yellow indicates Operating and Green indicates Completion. The sequence occurs three times, once for each syringe. The Control Module shall be in The Control Room.

THE DELIVERY MODULE

The Delivery Module consists of an eight inlet, one outlet stainless steel manifold containing two purge syringes filled with saline solution, two syringes filled with Sodium Pentathol, two syringes filled with Pancuronium Bromide, and two syringes filled with Potassium Chloride. The outlet is connected to a disposable intravenous administration set terminating in a needle tip and connected to a saline dispensing bag a short distance from the manifold. Additional hardware includes the cylinder matrix for supporting the syringe assembly, three electrical solenoids, three solenoid pull rods, six mechanical pull knobs, six connecting cables, six weight stop pins, six weighted pistons, six cylinders, and nine indicator lights. A total of eight disposable 60cc syringes are utilized in the system. During system makeready, all three solenoid pull rods are used, but only three of the six connecting cables, one for each set of two. The pull knobs are arranged in three pairs and both from each pair are pulled, one by each executioner, but only one from each pair is connected (either all odd or all even numbers). All manifold inlet connections are accomplished by luer lock fittings. Two purge syringe back flow stop brackets and twelve piston spaces are also part of The Delivery Module.

During makeready, the two saline syringes are used to bleed the system, the other syringes are installed, after filling, in the proper order, as assemblies, with the weighted piston, cylinder and piston stop pin. The cables are connected in the proper sequence and the solenoid pull rods inserted into the pistons. The Delivery Module shall be in the Execution Chamber.

PROCEDURE

It is suggested that the following procedure might be followed to facilitate a smooth execution. These dosages are established, although not recommended, through consultation with pharmaceutical manufacturers. We at Fred A. Leuchter Associates, Inc., not being pharmacologists, do not recommend, or in any way guarantee the efficacy of these chemicals or dosages, but simply communicate the recommendations of the manufacturers.

  1. Pre-injection 10cc antihistamine, one half hour prior to execution.
  2. Pre-injection 8cc 2% Sodium Pentathol (5 grams/250 ml, Abbott Labs #6108-01) five minutes prior to transmittal of subject to death chamber.
  3. Machine injection 15cc Sodium Pentathol 2 % Solution (as above) delivered over a ten second time period.
  4. One minute wait.
  5. Machine injection 15cc Pancuronium Bromide (Pavulon, Organon Drug Co., 2ml/2mg/ml) over a ten second time period.
  6. One minute wait.
  7. Machine injection 15cc Potassium Chloride (KCl Injectable solution).
  8. Two minute wait.
  9. Execution over.

SYSTEM NOTES

Times may be varied up to two and one half minutes by setting the timer in The Control Module.

Volumes to be dispensed may be varied by reducing the volumes in the syringe and utilizing the appropriate piston spacer.

Speeds of injection may be varied by using needles of different sizes. It is recommended not to exceed a twelve gauge. Fourteen gauge angiocaths are supplied.

The battery contains sufficient power at full charge to sustain operation for at least six usages of the system at fifteen minute time intervals before recharge is necessary. A uniform, slow-rate charge circuit, designed to guarantee a maximum battery life, will completely charge the battery over a 14 hour time period from 110 vac line power. The battery is a 12-volt 15-ampere hour battery.

DISCLAIMER

Fred A. Leuchter Associates, Inc. assumes no liability for the intended or actual use of this device.

OPERATIONAL PROCEDURE

PRE-SETUP

  1. Charge battery for 18 hours or more prior to usage. To charge battery, plug unit into 110 vac and turn key switch to charge position.
  2. Remove all cylinders and pistons. This is IMPORTANT.
  3. Test electrical operation. Turn both key switches to ARMED.
  4. Pull solenoid pins forward.
  5. Push both actuator switches and watch solenoid rods pull in sequence. One minute will occur between successive pulls. Depress each syringe switch located in Delivery Module after its solenoid rod has pulled.
  6. Repeat using override switches, twice: once for left and once for right position.
  7. Check syringe switches for proper clearance with piston disk in down position. Adjust switch up or down, with locknuts, for clearance at proper dosage. In all cases, switch must be set with piston resting on depressed syringe plunger (zero cc), with syringe properly installed in cylinder and on luer lock connection. If spacer is required for dosage, spacer must be affixed to piston. After height is determined, switch is locked into position with nuts; actuator arm at maximum position under switch disk, but not bent or flexed beyond proper depression point.
  8. Verify switch operation by manually dropping pistons on switches in sequence while repeating step 5.
  9. Recharge.

SETUP

  1. Remove all six pistons and cylinders from Delivery Module.
  2. Cylinders are numbered from right to left from 1 thru 6, with the two end syringes for purge.
  3. Consult chart for dosages and spacers.
  4. Assemble pistons and spacer with velcro pads.
  5. Fill purge syringes with 60cc saline solution and install on manifold by inserting into bracket and gently pushing luer lock tip into mating inlet with graduations on syringe facing operator. Turn one full turn to the right (clockwise from top).
  6. Depress both purge syringes until fluid escapes from all manifold inlets. Pinch outlet on extension set.
  7. Fill syringes with proper chemical dosages. Insert proper cylinder and lock proper syringe into cylinder retaining ring (slot) with graduations facing operator. Lift cylinder/syringe assemble and set luer lock tip of syringe into mating inlet. Pushing gently down, twist assemble one full turn to the right (clockwise from top) until graduations face operator. Install in order of 1 thru 6 from right. Install piston and stop pin after each cylinder/syringe installation. Note: cylinders 2, 4 and 6 use pistons with switch disk. Cylinders 1 & 2, Pentathol; cylinders 3 & 4, Pavulon; cylinders 5 & 6, KCl.
  8. Unpinch iv extension and depress both purge syringes until fluid escapes from outlet in iv extension. Install purge syringe back flow stop brackets. Pinch iv extension outlet.
  9. Slip solenoid pull rods through piston holes, cylinders 2, 4 & 6. Twist cylinder to check insertion. If it will not twist, insertion was proper. This is IMPORTANT.
  10. Install pull cable through piston holes cylinders 1, 3 & 5. Verify insertion visually. This is IMPORTANT. Use cables 1, 3 & 5 or 2, 4 & 6.STOP. Are the necessary spacers in place and backflow brackets installed? If so, proceed. When ready, remove piston stop pins. DANGER. System is now ARMED. Close door. See item 5 under use.
  11. Spacers and back flow stop brackets MUST be used as described.

USE

After completing test procedure and setup of Delivery Module, proceed as follows:

1. Pre-inject subject with 10cc antihistamine one half hour before execution.

2. Pre-inject subject with 8cc Pentathol 5 minutes before execution.

3. Strap subject to gurney and install iv, 1000ml 0.9 % sodium chloride via 14 gauge × 1.5 inch angiocath and start iv.

4. Transmit subject to execution chamber and connect Delivery Module iv to saline iv.

5. Remove piston stop pins from Delivery Module cylinders.

6. Turn both key switches to armed.

7. Begin execution on command by depressing both station buttons.

8. After activation, the following will occur:

  1. First syringe begins 4 seconds after start and a yellow light will light.
  2. Green light will indicate first stage complete in approximately 10 seconds.
  3. Second syringe will activate with yellow light one minute after first green light and will deliver solution and green light in approximately 10 seconds.
  4. Third syringe will activate with yellow light one minute after second green light and a third green light will indicate completion in approximately 10 seconds.
  5. After third green light, turn main switch to off. One minute later the execution will be completed.

(Approximate total time, 4 minutes.)

Note: if a malfunction occurs, it should be handled in the following manner:

  1. Any yellow light fails to operate.
    Cause: Timer malfunction.
    Remedy: Continue execution using electrical override switches. If the first yellow light fails, use all switches. If second yellow light fails, use switches two and three. If third yellow light fails, then use switch three.
  2. Any green light fails to operate.
    Cause: Defective syringe.
    Remedy: Continue execution using mechanical pull knobs. If the first green light fails, use all six knobs. If second green light fails, use phase two knobs. If third green light fails, then use only phase three knobs.

Note: Mechanical override can be used alone by pulling both knobs marked phase one and waiting 70 seconds, then pulling both knobs marked phase two and waiting 70 seconds, then pulling both knobs marked phase three.

9. After subject has been removed, remove back flow stop brackets from purge syringes and depress completely to purge system.

SIGNIFICANCE

Lethal injection is supposed to be more humane than other forms of execution. It may, therefore, have allowed executions to take place which would not otherwise have happened. Although the idea of putting someone to sleep, as one would an injured animal, may add some weight to the arguments of those supporting the death penalty, lethal injection is still associated with some serious drawbacks.

First, the condemned prisoner still undergoes the severe physical and mental stress associated with being on Death Row, especially as the execution time approaches. Second, the medicalization of execution gives rise to serious ethical and practical problems. Two of the three drugs used in lethal injection have genuine therapeutic uses. To use them to kill appears perverse and creates real ethical problems for doctors and nurses. Many professional medical associations, such as the American Medical Association, the World Medical Association, and the International Council of Nurses, have declared that taking part in executions is strictly against their ethical code. The fact that doctors and nurses are dedicated to healing the sick is a deeper issue than the misuse of the drugs used in lethal injection. To carry out tasks related to execution is a clear perversion of their duty, according to many professional organizations.

Yet lethal injection is a medical procedure and to be done correctly, some input is needed from medical personnel, argue prison authorities. This conflict has led to various legal actions being taken by doctors, their professional associations, and even prisoners against prison authorities in various states. In general, a doctor will at least certify that the prisoner is dead following lethal injection.

Although prison technicians can be trained in lethal injection, they are more likely to get it wrong than someone who is medically qualified. For instance, they may inject into a muscle instead of a vein, which can cause extreme pain, as can a clogged needle. Many prisoners have damaged veins because of previous intravenous drug use and it can be difficult to find a suitable vein for carrying out the lethal injection. This can lead to long and extremely stressful delays while the prisoner is still strapped in position, waiting for the procedure to begin.

The arguments against lethal injection are strengthened by a number of botched executions in recent years. In 1985, Stephen Peter Morin, who had a history of drug abuse, experienced a delay of forty-five minutes before his execution in Texas as technicians searched for a suitable vein. Also in Texas, the execution of Raymond Landry in 1988 was inter-rupted when the syringe came out of his vein, spraying the deadly chemicals all over the room. The syringe had to be re-inserted but this apparently took several minutes and was done without witness observation. Finally, the manual extract above refers to a lethal injection machine used in Missouri, the state where Bert Leroy Hunter suffered a botched execution in 2000. He had an unusual reaction to the drugs and one witness reported his violent convulsions during the injection process. Another Missouri execution went wrong in 1995, when the chemicals stopped circulating during the lethal injection of Emmitt Foster. The straps binding him to the gurney had been fixed too tight—the coroner had to step into the death chamber, diagnose the problem and order the straps to be loosened so the execution could proceed; death did not occur until half an hour after the process began.

FURTHER RESOURCES

Books

Web sites

Amnesty International. "Lethal Injection: The Medical Technology of Execution." 〈http://web.amnesty.org/library/Index/engACT500011998〉 (accessed February 26, 2006).

BBC News. "New Rules for U.S. Lethal Injection." 〈http://news.bbc.co.uk/2/hi/americas/4715034.stm〉 (accessed February 26, 2006).

Death Penalty Information Center. "Methods of Execution." 〈http://www.deathpenaltyinfo.org/article.php?scid=8&did=245〉 (accessed February 26, 2006).